Yokogawa EJA510E
In-Line Mount Absolute Pressure Transmitter
Pemancar tekanan absolut dan pengukur berperforma tinggi EJA510E dan EJA530E dilengkapi sensor resonansi silikon kristal tunggal dan cocok untuk mengukur tekanan cairan, gas, atau uap.
EJA510E dan EJA530E mengeluarkan sinyal DC 4 hingga 20 mA yang sesuai dengan tekanan yang diukur.
Ia juga dilengkapi respons cepat, pengaturan jarak jauh, dan pemantauan melalui komunikasi BRAIN atau HART dan diagnostik mandiri.
FOUNDATION Fieldbus, PROFIBUS PA, dan 1 hingga 5 V DC dengan protokol HART (Daya Rendah) juga tersedia.
Model seri EJA-E dalam konfigurasi standarnya, kecuali tipe Fieldbus, PROFIBUS, dan Daya Rendah, disertifikasi mematuhi SIL 2 untuk persyaratan keselamatan.
About OpreX™
OpreX™ is the comprehensive brand for Yokogawa’s industrial automation (IA) and control business and stands for excellence in the related technology and solutions.
It consists of categories and families under each category. This product belongs to the OpreX™ Field Instruments family that is aligned under the OpreX™ Measurement category.
Specifications
Measurement Types | |
---|---|
Primary Variable | Absolute Pressure |
Reference Accuracy | |
Primary Variable | ±0.055% of Span ±0.04% of Span (with /HAC option) |
Stability (All Normal Operating Conditions) | |
Primary Variable | ±0.2% of URL per 10 years |
Response Time | |
Primary Variable | 90ms |
Rangeability | |
Primary Variable | A capsule 20:1 B, C and D capsules: 100:1 |
Burst Pressure (Absolute) | |
EJA510E | A, B, and C Capsules: 4,350 psi (30 MPa) D Capsule: 19,100 psi (132 MPa) |
Specification Conformance | |
EJX-A Series | ±3σ |
Span and Range Limits
(For EJA510E, values are in absolute pressure and lower range limits are 0.)
Measurement
Span/Range |
MPa | psi (/D1) | bar (/D3) | kgf/cm2
(/D4) |
|
A |
Span | 10 to 200
kPa |
1.45 to 29 | 0.1 to 2 | 0.1 to 2 |
Range | -100 to 200 kPa | -14.5 to
29 |
-1 to 2 | -1 to 2 | |
B |
Span | 0.02 to 2 | 2.9 to 290 | 0.2 to 20 | 0.2 to 20 |
Range | -0.1 to 2 | -14.5 to
290 |
-1 to 20 | -1 to 20 | |
C |
Span | 0.1 to 10 | 14.5 to
1450 |
1 to 100 | 1 to 100 |
Range | -0.1 to 10 | -14.5 to
1450 |
-1 to 100 | -1 to 100 | |
D |
Span * | 0.5 to 50 | 72.5 to
7200 |
5 to 500 | 5 to 500 |
Range * | -0.1 to 50 | -14.5 to
7200 |
-1 to 500 | -1 to 500 |
*: Maximum value shall be 70 MPa, 10150 psi, 700 bar or 700 kgf/cm2 respectively when /HG is specified.
Reference Accuracy of Calibrated Span
(includes the effects of terminal-based linearity, hysteresis, and repeatability)
Measurement
span |
Reference Accuracy | |
Span≥X | Span<X | |
A | ±0.055%
of Span |
±(0.0055 URL/ span)% of Span |
B | ||
C | ||
D | ±(0.0088 ∗50 MPa/ span)%
of Span |
[When /HAC is specified]
Measurement
span |
Reference Accuracy | |
Span≥X | Span<X | |
A | ±0.04% of Span | ±(0.004 URL/ span)% of Span |
B | ±(0.005+0.0035 URL/ span) %
of Span |
|
C | ||
D | ±(0.0064 ∗50 MPa/ span)%
of Span |
Measurement
span |
A | B | C | D |
X | 20 kPa (2.9 psi) | 0.2 MPa
(29 psi) |
1 MPa (145 psi) | 8 MPa (1160 psi) |
URL
(Upper range limit) |
200 kPa
(29 psi) |
2 MPa (290 psi) | 10 MPa (1450 psi) | 50 MPa (7200 psi) |